Hit Enter to Search
Precont®  DD109B

Precont® DD109B

Differential pressure transmitter For monitoring the differential pressure of air or other non-flammable and non-aggressive gases

Inquire now

Advantages
  • Precise detection of differential pressures in dry non-aggressive gases
  • Suitable for dusty environments (ATEX approval)
  • Ideal for filter monitoring
  • Optimal adjustment of the measuring range
  • Adjustment of the zero point possible
  • Fast commissioning due to easy installation and easily accessible connection area
Application description

The differential pressure measuring transducer Precont DD109B is a universal measuring transducer for small and medium pressures. By fitting it with various pressure sensors, measuring ranges between 2.5 and 100 mbar can be realised. Two connection cables are used for the power supply. The feed current is the measurement signal from 4...20 mA. The operating state is indicated with an LED.

Technichal data

Measuring range (mbar) 0-10; 0-25; 0-50; 0-100; 0-1000
Max. Differential pressure Measuring range 0-10 mbar: 100 mbar / Measuring range 0-100 mbar: 750 mbar / Measuring range 0-1000 mbar: 4000 mbar
Medium air, as well as dry, non-aggressive gases
Basic accuracy ± 1% of the final value
Temperature drift ± 0.05% / K of the final value
Hysteresis ± 0.5% of full scale
Measuring system semiconductor sensor
Power supply Ub = 10 ... 36 V DC
Analog output 4 ... 20 mA, 2-wire technology
Max. Permissible load RA ≤ (Ub - 9 V) / 0.02 A
Pressure connection bulkhead push-in fittings for 6 mm / Tube outer diameter
Housing Dust-proof polycarbonate housing
Dimension 113 × 80 × 60 (W × H × D)
Protection class IP 65
Mounting wall mounting, installation position vertical
Connection Spring-loaded terminals
Cable gland 1 × M 16 × 1.5 N for cable diameters of 4 ... 8 mm
Device Identification / Explosion Protection II3D Ex tc IIIC T135 ° Dc / II3G Ex nR IIC T4 Gc

Climatic environmental conditions
Storage
-20 ° ... + 70 ° C
Transport -40 ° ... + 85 ° C
Operation -20 ° ... + 55 ° C
Relative humidity 75% rel. Humidity, no condensation

Safety metrics
MTBF 646 years
MTTFd 1292 years
Service life Max. 10 years
Category according to EN-ISO 13849 B
Calculation method according to ENISO 13849 Parts-Count
Environmental temperature 50°C