Interchangeable pressure board with solenoid valve and a terminal for K thermocouple probe for C 310 and CA 310 sensors.
Inquire nowRef: SPI2-100 / SPI2-500 / SPI2-1000 / SPI2-10000
The SPI2 pressure boards for C 310 and CA 310 sensors have a temperature compensation of the gain from 0 to 50°C and an auto-calibration process which guarantee an excellent stability and a perfect reliability of the measurement as much in low as high range over time.
Auto-calibration principle: the microchip of the sensor manages a solenoid valve which compensates the drift of the sensitive element over time. The zero-point adjustment is ensured by this compensation. The differential pressure measurement is then independent from the environmental conditions of the sensor.
Solenoid valve lifetime: 100 million of cycles
Advantage: no zero-point drift
Periodicity of the auto-calibration: from 1 to 60 minutes, or deactivated.
Measure details | |
Units | °F, °C, Pa, kPa, mbar, hPa, inWg, mmH?O, mmHg, daPa |
Measuring ranges | |
SPI2-100 | -100.00 Pa ... 500.00 Pa |
SPI2-500 | -500.00 Pa ... 500.00 Pa |
SPI2-1000 | -1000.00 Pa ... 1000.00 Pa |
SPI2-10000 | -10000.00 Pa ... 10000.00 Pa |
SPI2-100 / SPI2-500 / SPI2-1000 / SPI2-10000 | -200.00 °C ... 1300.00 °C |
Technical details | |
Parameters | Pressure Temperature |
Standards | UKCA |
Wireless connection | No |
2 Pressure connectors
2 Silicone tubes
1 Adjustment certificate